1. Select Time Scan from the Main menu.
2. Set the operating conditions as follows:
Data
Mode
ABS
Wavelength
(nm)
500
Scan
time
(s)
3600
UP
Scale
0.001
LO
Scale
-0.001
Lamp Change Mode
Auto
Lamp Change WL (nm)
340.0
WI Lamp
ON
D2
Lamp
ON
Response
Medium
3. After setting up the conditions, press 0 to display the measurement screen.
4. Verify that the Sample and Reference cuvette holders are empty. And then
press Auto Zero.
5. Press START to start a measurement.
6. After measurement, check the baseline stability as shown figure below:
D
3600sec
4.2.4 Confirmation of Baseline Flatness
Be sure to warm up the instrument sufficiently (approx. 2 hours), and then
proceed as instructed below:
1. Select Wavelength Scan from the Main menu.
2. Set the operating conditions as follows:
Data Mode
ABS
Start
WL
(nm)
950
Stop
WL
(nm)
200
Scan
Speed
(nm/min)
400
UP
Scale
0.005
LO
Scale
-0.005
Lamp Change Mode
Auto
Lamp Change WL (nm)
340.0
WI Lamp
ON
4-3