Parameter Setup
1
Lamp Change Mode
Auto
2
Sample Setup
2
Lamp Change WL ( nm )
350.0
3
Curve Setup
3
WI Lamp
On
4
Print Setup
4
D2 Lamp
Off
5
System Setup
6
Save Parameter
0
Forward
Select Function: _
Lamp Change Mode:
The followings three Lamp Change Mode are available.
Auto: Changeover is automatically made between D2 Lamp and WI
Lamp.
D2 Lamp Only: Using D2 lamp over the entire wavelength region.
WI Lamp Only: Using WI lamp over the entire wavelength region.
Lamp Change WL (nm):
This parameter is used for selection of light source switching
wavelength.
This spectrophotometer is provided with WI lamp for visible range
measurement and D2 lamp for ultraviolet range measurement. The
wavelength is selectable with a range of 325.0 ~ 370.0 nm in increments
of 0.1 nm.
WI Lamp: Turn on or turn off the WI lamp.
D2 Lamp: Turn on or turn off the D2 lamp.
Note: If turn on the lamp but display “OFF”, perhaps the lamp is error.
3.3.1.6 Save Parameter
3-11