5.10 Settings sensor
5.10.1 Sensor change
With this function, a sensor can be changed during operation
without triggering a fault signal at the central controller. It also
ensures that all sensor data in the microprocessor can be
saved in the sensor memory (EEPROM) before disconnecting
the sensor.
A sensor can be changed at any time. However, for safety
reasons a fault alarm is generated when the sensor is
disconnected until a new sensor is connected.
1. Select
Settings > Sensor > Sensor change
and confirm.
The maintenance signal is generated on the analog
interface, the following display shows: “Please remove the
sensor”.
2. Replace the old sensor with a new sensor, see Section 6.4
When the new sensor is installed:
Display: “Loading data, please wait”.
When the sensor data are loaded:
Display: “Data loaded”.
3. Select
Back to menu
and confirm.
The maintenance signal on the analog interface remains
pending until the sensor is warmed-up.
The warm-up time depends on the type of sensor and its
history. For example, if the sensor has already warmed-up on
another instrument and the operation was only interrupted for
a short time, the warm-up time can be shorter than the time
specified in the Instructions for Use for the sensor.
Maximum warm-up time of a new sensor: see the Instructions
for Use for the sensor or Section 10.1 on Page 38 for Oxygen:
EN 50104 approvals.
5.10.2 Auto CAL
1. Select
Settings > Sensor > auto calibration
and confirm.
2. Select
Enable
or
Disable
and confirm.
5.10.3 Sensor test
This function is only active for Polytron 8100 using the EC
sensor.
This function is only active if the sensor test or diagnostics
dongle is installed.
This function activates the sensor test. The instrument
routinely checks the sensor for proper function. If the sensor
does not pass the self-test, a warning or fault message will be
issued.
1. Select
Settings > Sensor > Sensor test
and confirm.
2. Select
Enable
or
Disable
and confirm.
5.10.4 Sensor lock
This function activates the sensor lock feature.
1. Select
Settings > Sensor > Sensor lock
and confirm.
2. Select
On
or
Off
and confirm.
5.10.5 Sensor type
This function is only active for Polytron 8200 or Polytron 8310
using the DD, LC or DSIR sensor.
NOTICE
If a sensor of the same type (same part number) was
previously installed, the instrument specific
configuration is retained (gas type, measurement
range, test gas, calibration interval, etc.)
Otherwise the factory default settings of the new
sensor (see Instructions for Use for the sensor) are
uploaded and will overwrite the instrument specific
configuration.
This can be prohibited if the sensor lock function (see
Section 5.10.4 on Page 27) is activated.
NOTICE
This function is only active if enabled here.
i
i
i
i
CAUTION
If
turned
On
the instrument will reject any sensor
whose part number does not match the
part number of the previously installed
sensor.
If
turned
Off
the instrument will accept any suitable
sensor, uploading the default settings of
this sensor. However, this will overwrite
any customized settings with the default
settings of the new sensor.
!
Содержание Polytron 8100
Страница 2: ......
Страница 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Страница 4: ...1 2 3 4 1 2 1 2 3 4 1 2 VDC mA VDC mA F G I H J K L t s 1 1 mA 0 7 5 3 t s T 1 4 20 mA T 2 3 SAMPLE...
Страница 172: ...172 Dr ger Polytron 8100 1 1 1 Dr ger Dr ger Dr ger 6 196 Dr ger EN IEC 60079 1 II 1 5 5 1 5 6 1 5 7 2014 34 EC 1 2 i i...
Страница 174: ...174 Dr ger Polytron 8100 3 3 1 10 1 204 Hawke A501 421 A NPT NPT 10 9 210 20 7 12 4 8 55 C 25 C 5 7 3 174...
Страница 182: ...182 Dr ger Polytron 8100 4 5 1 4 3 OK OK OK OK 5 8 2 188 _ _ _ 1 _ _ _ 2 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 198: ...198 Dr ger Polytron 8100 6 2 3 1 2 3 4 5 6 1 2 3 4 5 6 1 2 3 4 5 OK 6 6 1 1 196 6 1 2 196 Dr ger H2S ppm 25 i i 15 i i...
Страница 201: ...Dr ger Polytron 8100 201 6 4 B 1 2 2 2 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 6 2 196 10 SIL 5 6 185 5 10 4 193 i i i i...
Страница 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 239: ...Dr ger Polytron 8100 239 6 4 B 1 2 2 2 mm 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 235 6 2 10 SIL 225 5 6 233 5 10 4 i i i i...
Страница 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Страница 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Страница 253: ......