18
Dräger Polytron 8100
Menu
5.2
Information instrument
5.2.1
Warning messages
This function displays warning messages in plain text and the
corresponding number see Section 6.3.2 on Page 34. The [ ! ]
symbol is displayed if there are warning messages available.
1. Select
Information > Instrument > Warnings
and
confirm.
If there are several warning messages available, there is
an indicator in the upper right-hand corner of the display,
e.g. 1/3 = Screen 1 of 3.
5.2.2
Fault messages
This function displays fault messages in plain text and the
corresponding number see Section 6.3.1 on Page 33. The [ X ]
symbol is displayed if there are fault messages available.
1. Select
Information > Instrument > Faults
and confirm.
2. If there are several fault messages available, there is an
indicator in the upper right-hand corner of the display e.g.
1/2 = Screen 1 of 2.
5.2.3
Installed modules
This function displays an overview of the installed hardware
modules.
1. Select
Information > Instrument > Modules
and confirm.
The list contains modules. The installed modules are
identified by the
symbol. The
symbol means that the
module is not installed.
2. Highlight a module and access detailed information with
[OK].
5.3
Information sensor
This function contains information on calibration dates.
5.3.1
Display last calibration date
1. Select
Information > Sensor > Last cal. date
and
confirm. The calibration date, unit, gas and concentration
of the last calibration are displayed.
5.3.2
Display next calibration due date
1. Select
Information > Sensor > Next cal. date
and
confirm.
5.3.3
Vitality
1. Select
Information > Sensor > Vitality
and confirm. The
sensor vitality is displayed in %.
5.3.4
Sensor temperature
1. Select
Information > Sensor > Sensor temp.
and
confirm. The current and the highest sensor temperature is
displayed.
5.4
Information data-logger
These functions contain information about the data-logger.
5.4.1
Display the data-logger status
1. Select
Information > Data logger > Logger status
and
confirm.
2. To change the status see Section 5.11 on Page 29.
5.4.2
Display graph
Display the history of the past 15-minutes on a time/
concentration graph.
1. Select
Information > Data logger > Graph
and confirm.
5.5
Calibration
5.5.1
Zero calibration
Allows adjusting the zero reference point of the sensor, see
Section 6.2.1 on Page 31.
5.5.2
Span calibration
Allows adjusting the sensitivity to match the known
concentration of an applied calibration gas, see Section 6.2.2
on Page 31.
5.5.3
Auto calibration
Automated sequence reducing the number of steps during
calibration; see Section 6.2.3 on Page 32.
Содержание Polytron 8100
Страница 2: ......
Страница 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Страница 4: ...1 2 3 4 1 2 1 2 3 4 1 2 VDC mA VDC mA F G I H J K L t s 1 1 mA 0 7 5 3 t s T 1 4 20 mA T 2 3 SAMPLE...
Страница 172: ...172 Dr ger Polytron 8100 1 1 1 Dr ger Dr ger Dr ger 6 196 Dr ger EN IEC 60079 1 II 1 5 5 1 5 6 1 5 7 2014 34 EC 1 2 i i...
Страница 174: ...174 Dr ger Polytron 8100 3 3 1 10 1 204 Hawke A501 421 A NPT NPT 10 9 210 20 7 12 4 8 55 C 25 C 5 7 3 174...
Страница 182: ...182 Dr ger Polytron 8100 4 5 1 4 3 OK OK OK OK 5 8 2 188 _ _ _ 1 _ _ _ 2 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 198: ...198 Dr ger Polytron 8100 6 2 3 1 2 3 4 5 6 1 2 3 4 5 6 1 2 3 4 5 OK 6 6 1 1 196 6 1 2 196 Dr ger H2S ppm 25 i i 15 i i...
Страница 201: ...Dr ger Polytron 8100 201 6 4 B 1 2 2 2 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 6 2 196 10 SIL 5 6 185 5 10 4 193 i i i i...
Страница 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 239: ...Dr ger Polytron 8100 239 6 4 B 1 2 2 2 mm 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 235 6 2 10 SIL 225 5 6 233 5 10 4 i i i i...
Страница 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Страница 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Страница 253: ......