
Maintenance
Dräger Polytron 8100
31
6.2.1
Zero Calibration
1. Select
Calibration > Zero Calibration
and confirm.
The Maintenance signal is transmitted and the symbol [ ]
is displayed.
The message
Apply zero gas
is displayed.
2. Apply Synthetic Air or Nitrogen.
3. Select
Next
and confirm.
The current value is displayed.
After the displayed value is stable:
1. Select
Calibrate
and confirm.
The message
Please wait...
is displayed.
The new current value is displayed.
2. Select
Next
and confirm.
3. Turn off gas flow and remove the calibration adapter from
the sensor or disconnect tubing.
If the current value is not within the alarm range:
4. Select
Next
and confirm.
The instrument returns to the calibration menu.
6.2.2
Span calibration
1. Select
Calibration > Span calibration
and confirm.
The Maintenance signal is transmitted and the symbol [ ]
is displayed.
The parameters for the calibration gas are displayed, e.g.:
Depending on the sensor these parameters can be changed:
1. Select
Gas
and confirm.
2. Select the calibration gas from the list and confirm.
3. Select
Unit
and confirm.
4. Select the unit of measurement from the list and confirm.
5. Select
Conc.
and confirm.
6. Set the concentration of the calibration gas.
If the settings are correct:
1. Select
Next
and confirm.
A message like
Gas flow ON H2S
is displayed.
2. Apply calibration gas.
3. Select
Next
and confirm.
The current value is displayed.
After the displayed value is stable:
1. Select
Next
and confirm.
The message
Please wait...
is displayed.
The new current value is displayed.
2. Select
Next
and confirm.
3. Turn off gas flow and remove the calibration adapter from
the sensor or disconnect tubing.
If the current value is not within the alarm range:
4. Select
Next
and confirm.
The instrument returns to the calibration menu.
A calibration can be aborted at any time.
WARNING
Ambient air can be used to zero the sensor instead of
Nitrogen or Synthetic Air only if the area is known to be
free of the target gas or any gas to which the sensor
may be cross-sensitive (as listed on the sensor data
sheet). In this case, no cylinder or calibration adapter
is needed for the zero calibration.
NOTICE
For Oxygen (O
2
) sensors a zero calibration does not
change any value in the firmware or sensor. Thus, the
zero calibration function can be used to check the
correct zero reading when Nitrogen is applied to the
sensor. The Polytron 8100 will indicate a fault state if a
zero check fails with a deviation of more than 0.6 Vol%
O
2
from zero. In case of a fault, repeat the zero check
or change the sensor if necessary. Use Nitrogen (N
2
)
for verification.
!
i
i
Gas
H2S
Unit
ppm
Conc.
25
Содержание Polytron 8100
Страница 2: ......
Страница 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Страница 4: ...1 2 3 4 1 2 1 2 3 4 1 2 VDC mA VDC mA F G I H J K L t s 1 1 mA 0 7 5 3 t s T 1 4 20 mA T 2 3 SAMPLE...
Страница 172: ...172 Dr ger Polytron 8100 1 1 1 Dr ger Dr ger Dr ger 6 196 Dr ger EN IEC 60079 1 II 1 5 5 1 5 6 1 5 7 2014 34 EC 1 2 i i...
Страница 174: ...174 Dr ger Polytron 8100 3 3 1 10 1 204 Hawke A501 421 A NPT NPT 10 9 210 20 7 12 4 8 55 C 25 C 5 7 3 174...
Страница 182: ...182 Dr ger Polytron 8100 4 5 1 4 3 OK OK OK OK 5 8 2 188 _ _ _ 1 _ _ _ 2 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 198: ...198 Dr ger Polytron 8100 6 2 3 1 2 3 4 5 6 1 2 3 4 5 6 1 2 3 4 5 OK 6 6 1 1 196 6 1 2 196 Dr ger H2S ppm 25 i i 15 i i...
Страница 201: ...Dr ger Polytron 8100 201 6 4 B 1 2 2 2 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 6 2 196 10 SIL 5 6 185 5 10 4 193 i i i i...
Страница 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 239: ...Dr ger Polytron 8100 239 6 4 B 1 2 2 2 mm 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 235 6 2 10 SIL 225 5 6 233 5 10 4 i i i i...
Страница 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Страница 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Страница 253: ......