Dräger Polytron 8100
Menu
Set concentration
This function sets the current corresponding to any
concentration value between 0 and full-scale.
1. Select
Settings > Communication > Analog interface >
Set concentr.
and confirm.
2. Once the alarms at the central controller are inhibited,
confirm the message
Inhibit all alarms
.
3. Select the line for editing the concentration and confirm.
4. Set the concentration and confirm.
The setting for the concentration is displayed.
5. Select
Next
and confirm.
6. Select
Set concentr. out
or
Concentration off
and
confirm.
7. Select
Next
and confirm.
The function is terminated.
8. Once the alarms at the central controller are enabled
again, confirm the message
Enable all alarms
.
Set fault
This function sets the current to the fault current.
1. Select
Settings > Communication > Analog interface >
Set fault
and confirm.
2. Select
Enable
or
Disable
and confirm.
Set warning
This function sets the current to the warning current.
1. Select
Settings > Communication > Analog interface >
Set warning
and confirm.
2. Select
Enable
or
Disable
and confirm.
Set maintenance
This function sets the current to the maintenance current.
1. Select
Settings > Communication > Analog interface >
Set mainten.
and confirm.
2. Select
Enable
or
Disable
and confirm.
Set beam block signal
This function is only available for Polytron 87x0 (using the
sensor PIR 7x00).
5.9.3
Modbus
For further information regarding the configuration of the
modbus interface consult the Technical Modbus Handbook,
5.9.4
Profibus address
For further information regarding the configuration of the
Profibus PA or Foundation Fieldbus H1 interface consult:
- Technical Profibus Handbook or
- Technical Fieldbus Handbook
NOTICE
Using the fieldbus interface no relays are available.
The fieldbus interface is not in accordance with the
BVS 13 ATEX G 001 X.
NOTICE
Using the fieldbus interface no relays are available.
The fieldbus interface is not in accordance with the
BVS 13 ATEX G 001 X.
i
i
i
i
Содержание Polytron 8100
Страница 2: ......
Страница 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Страница 4: ...1 2 3 4 1 2 1 2 3 4 1 2 VDC mA VDC mA F G I H J K L t s 1 1 mA 0 7 5 3 t s T 1 4 20 mA T 2 3 SAMPLE...
Страница 172: ...172 Dr ger Polytron 8100 1 1 1 Dr ger Dr ger Dr ger 6 196 Dr ger EN IEC 60079 1 II 1 5 5 1 5 6 1 5 7 2014 34 EC 1 2 i i...
Страница 174: ...174 Dr ger Polytron 8100 3 3 1 10 1 204 Hawke A501 421 A NPT NPT 10 9 210 20 7 12 4 8 55 C 25 C 5 7 3 174...
Страница 182: ...182 Dr ger Polytron 8100 4 5 1 4 3 OK OK OK OK 5 8 2 188 _ _ _ 1 _ _ _ 2 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 198: ...198 Dr ger Polytron 8100 6 2 3 1 2 3 4 5 6 1 2 3 4 5 6 1 2 3 4 5 OK 6 6 1 1 196 6 1 2 196 Dr ger H2S ppm 25 i i 15 i i...
Страница 201: ...Dr ger Polytron 8100 201 6 4 B 1 2 2 2 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 6 2 196 10 SIL 5 6 185 5 10 4 193 i i i i...
Страница 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Страница 239: ...Dr ger Polytron 8100 239 6 4 B 1 2 2 2 mm 3 3 4 5 4 Dr ger 5 6 7 2 22 8 9 235 6 2 10 SIL 225 5 6 233 5 10 4 i i i i...
Страница 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Страница 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Страница 253: ......