Trouble Shooting
Message in t he
Cause
Explanation/Remedy
alarm window
No communication
Device defective.
Contact your local DrägerService
control- / charge-board
System leakage
Leak in ventilation hose system and/or test lung.
Check hoses, breathing valve,
flow sensor and test lung for leaks and
replace if necessary.
No test lung
Test lung not connected or major leakage.
Connect test lung.
Check hoses, breathing valve, flow sensor
and test lung for leaks and replace if necessary.
Breathing valve inop
Breathing valve has malfunctioned.
Check correct condition of breathing valve
including diaphragm and rubber disc; fit a new
breathing valve if necessary or use a new
disposable hose set.
Pressure
The ventilation hose system has not been
Connect ventilation system correctly.
measurement inop
connected correctly.
Pressure measurement is not possible.
Contact your local DrägerService for additional
support.
PEEP-valve inop
Internal leak in system.
Check hoses, breathing valve, flow sensor and
test lung for leaks and replace if necessary.
Device defective.
Contact your local DrägerService for additional
support.