Trouble Shooting
Messages in the alarm window
Alarm
Cause
Remedy
!!! Apnea
Spontaneous breathing by the patient
Ventilate in VC-CMV mode.
has failed, or disconnection.
Ensure that hose connections are tight.
Faulty flow sensor.
Replace flow sensor.
!!! Apnea ventilation
The ventilator has automatically
Check ventilation mode. To return to
(only for SpnCPAP)
switched over to mandatory ventilation
original ventilation mode: Press
after detecting an apnea
the
Alarm Reset
key.
(only in SpnCPAP mode).
!! Charge int. battery
Oxylog 2000
plus
draws its power from
The ventilator must immediately
the internal battery due to the absence
be reconnected to the mains supply,
of an external DC supply. Only a few
an onboard DC supply or a fully
minutes of operating time remain
charged battery.
(approximately 10 minutes).
!! Check settings flow
The flow resulting from the settings for
Change tidal volume
VT
or inspiratory
"Tidal volume
VT
per unit time" is not
time
Ti
or ventilation time ratio
I:E
.
possible.
!! Check settings time
The expiration time resulting from the
Change
RR
or
I:E
or
Ti
.
settings for
RR
and
I:E
or
Ti
is not possible.