![Dräger Oxylog 2000 plus Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/drager/oxylog-2000-plus/oxylog-2000-plus_pocket-manual_2526811032.webp)
32
Operation / Settings
VC-SIMV (optional PS)
VC-SIMV
-
V
olume
C
ontrolled -
S
ynchronized
I
ntermittent
M
andatory
V
entilation
For patients with inadequate spontaneous breathing, or for patients who
are to be weaned gradually.
Fixed mandatory minute volume MV is set with tidal volume VT and
ventilation respiratory rate RR. The patient can breathe spontaneously
between the mandatory ventilation strokes and thus contribute to the
total minute volume. Spontaneous breathing can be assisted with PS.
Set the ventilation pattern with the controls below the display:
• Tidal volume
VT
.
• Respiratory Rate
RR
.
(minimum possible respiratory rate: 2 per min).
• Maximum airway pressure
Pmax
.
• O
2
setting
FiO
2
.
• Inspiration time
Ti
.
• Plateau time
Tplat
%, in % of the inspiration time.
• Positive end expiratory pressure
PEEP
• Sensitivity
Trigger
.
Paw
Flow
T
i
P
max
PEEP
rapid
rise time
slow
rise time
Pressure
support PS
Trigger window
RR
Insp. Flow
t
t
1
Содержание Oxylog 2000 plus
Страница 2: ......
Страница 9: ...9 A Filter cartridge for intake of ambient air B Rating plate A B Rear view ...
Страница 51: ...51 Notes ...