45
Trouble Shooting
Messages in the alarm window
Alarm
Cause
Remedy
!!! Leakage (not in NIV)
The measured expiratory tidal volume VTe
Repair leaks in hose system and possibly in
is approximately 40% lower than the
the tube.
inspiratory value.
Use new flow measuring hoses.
Faulty flow sensor.
Replace flow sensor.
The ventilator may not function properly.
Contact your local DrägerService
for additional support.
!! Loss of data
Technical defect.
Contact your local DrägerService for
additional support.
!! Loudspeaker inop
Technical defect.
Contact your local DrägerService for
additional support – restricted operation
is now possible.
!!! MV high
The upper alarm limit for the minute
Check patient's condition, check
volume MV has been exceeded.
ventilation pattern, adjust alarm
limits if necessary.
Faulty flow sensor.
Replace flow sensor.
The ventilator may not function properly.
Contact your local DrägerService
for additional support.