Trouble Shooting
Messages in the alarm window
Alarm
Cause
Remedy
!!! Paw high
The alarm limit Pmax for the airway pressure
Check patient's condition, check
has been reached. Patient "fights" the
ventilation pattern, adjust alarm
ventilator, coughing.
limits if necessary.
Ventilation hose kinked, or obstructed.
Check hose system, breathing
valve and tube.
!!! Paw low
No pressure difference >5 mbar between
Inflate cuff and check for leaks.
inspiration and expiration or set pressure
level is not achieved.
Leak in cuff.
Leakage or disconnection.
Check hose system for leaking
connections. Ensure that the
breathing valve has been installed
correctly.