Messages in the alarm window
Alarm
Cause
Remedy
!!! Confirm device OFF
Key
has been pressed for 3 seconds.
To switch OFF: confirm with the rotary knob.
wit h rotary knob
To continue ventilation, press key
again.
!!! Device failure
Technical defect.
Contact your local DrägerService
for additional support.
!! Flow measurement inop
Measurement hoses for flow measurement
Ensure flow measurement hoses
hoses kinked, disconnected or leaking.
are connected correctly.
Flow sensor defective.
Replace flow sensor.
Technical defect.
Contact your local DrägerService
for additional support – restricted
operation is now possible.
!! Gas delivery failure
Technical defect.
Contact your local DrägerService
for additional support.