![Dräger Oxylog 2000 plus Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/drager/oxylog-2000-plus/oxylog-2000-plus_pocket-manual_2526811036.webp)
36
Operation / Settings
NIV – Non-Invasive Ventilation (optional)
NIV (optional) can only be activated as a supplementary function in the
pressure controlled ventilation modes SpnCPAP and SpnCPAP / PS.
Mask leakages are detected by the device, compensated and included
in the measured values for VTe and MVe. The leakage alarm is inactive.
• The supplement NIV appears in the upper section of the display.
Please refer to the NIV section of the Instructions for Use for a clear
understanding of risks associated with NIV use.