Trouble Shooting
Messages in the alarm window
Alarm
Cause
Remedy
!! Paw measurement inop
Fault in flow measurement hoses.
Ensure hose system for loose connections.
Ensure flow measurement hoses are
connected correctly.
Technical defect.
Contact your local DrägerService for
additional support – restricted operation
is now possible.
! Self test OK
The device has been switched on and
The message can be confirmed or it will be
the selftest completed successfully.
cancelled automatically with the next message.
!! Set correct FiO
2
The control knob for setting
O
2
AirMix
Set the control knob in the right
or 100%
O
2
is set in a middle position.
or left position.
! Settings not confirmed
Parameters have been changed on the
Press the rotary knob to confirm the
screen but not confirmed.
parameter changes.
!! Supply pressure low
Supply pressure <270 kPa.
Ensure that supply pressure exceeds 270 kPa.