Operation / Settings
Oxylo
g 2000
plus
A
Trigger (VC-AC)
Activating/setting t he trigger
1
Press the key
Settings
(A) until the trigger parameter is
displayed.
2
Select the line
Trigger
on the display and then set and confirm
the value with the rotary knob.
Small value = high sensitivity.
The ventilation mode
VC-AC
is shown on the display.
Deactivate trigger
1
Set a value less than 3 L/min or greater than 15 L/min (
off
is
displayed instead of a value).
2
Press the rotary knob to confirm.
The last effective trigger value is adopted by the ventilator when
changing from VC-AC to SpnCPAP.
Successful patient triggering is briefly indicated by an asterisk (*)
in the middle of the status and alarm message window.
NOTE
If in VC-CMV the trigger is set »
on
«, the ventilation mode changes into VC-AC.