20
Preparation
Connecting the gas supply from an O
2
cylinder
Connecting t he gas supply supply from an O
2
cylinder
1
Use a full O
2
cylinder.
2
Connect the pressure reducer (270 to 600 kPa delivery pressure,
500 kPa nominal pressure) to the O
2
cylinder.
3
Connect the O
2
medical gas hose (A) to the
Oxylog 2000
plus
.
4
Connect the O
2
medical gas hose to the pressure reducer (B).
5
Rotate the cylinder valve (C) slowly and open fully.
O
2
A
B
C