26
Operation / Settings
Connect t he test lung
1
Connect the angled adapter (A) to
the breathing valve.
2
Connect the catheter connector
(B), diameter 7 mm, to the angled
adapter. The catheter connector
simulates the resistance of the
airways.
3
Connect the test lung (C).
Switch ON
1
To switch ON briefly press
the
key (B).
The device performs a self-test and
the operator is prompted, on the
display, to activate the configuration
menu or device check:
2
Press the rotary knob (A) to
confirm, before the bar is full.
3
Select Device check in the main
menu and confirm.
A
C
B
Perform device check
Oxylog 2000
plus
A
B
NOTE:
The device check can be
discontinued at any time by
pressing the »
Alarm Reset
« key.