76
Routine Maintenance
Calibration Using the Flow-Match Regulator
Connecting the Flow-Match Regulator to the Cylinder
To connect the flow-match regulator to the gas cylinder:
1. Connect the regulator to the calibration gas cylinder.
2. If you are using a portable tank of calibration gas (Part No.
68 10 687), connect the regulator (Part No. 68 10 688) directly to
the tank.
3. When the regulator is connected properly, you can read the cylin-
der contents from the regulator gauge.
4. Connect the adapter tubing to the regulator.
5. Check if the Multi-PID 2 is sucking in the gas pressure free, e.g.
by means of a T-piece; paying attention the fact that the flow from
the gas cylinder is higher than the inlet flow of the Multi-PID 2.
Figure 18 – Calibration with Flow-Math
WARNING:
Observe proper handling procedure for all gases!
0
1
623958_0
1
_en.eps
CAL
Multi-PID 2
ALARM
CHARGE
cancel new next
Calibration
Calibration gas
concentration
100.0
Содержание Multi-PID 2
Страница 7: ...7 Notices and Warnings Notices and Warnings...
Страница 14: ...14 Notices and Warnings...
Страница 15: ...15 Introduction Introduction...
Страница 25: ...25 Using the Multi PID 2 Using the Multi PID 2...
Страница 42: ...42 Using the Multi PID 2...
Страница 43: ...43 Connecting Accessories Connecting Accessories...
Страница 52: ...52 Connecting Accessories...
Страница 53: ...53 Menu Functions Menu Functions...
Страница 71: ...71 Routine Maintenance Routine Maintenance...
Страница 86: ...86 Routine Maintenance...
Страница 87: ...87 Troubleshooting Troubleshooting...
Страница 93: ...93 Appendices Appendices...
Страница 104: ...104 Appendices...