90
Troubleshooting
Pump Error
If the pump sounds labored,
the pump is operating out-
side its normal operating
range.
Check for obstructions in the sample
line.
Verify that the sample line, sample
probe and inlet filter are not blocked.
Ensure that the sample outlet is not
blocked.
The Multi-PID 2 has aspirated
a liquid.
Contact Dräger Safety.
The pump has failed.
Contact Dräger Safety.
Blocked Filter
Replace the inlet filter Dräger Safety
Part No. 83 17 681.
Instrument over
range
High concentrations of gases
will saturate the detector and
electronics.
Wait until the display returns to nor-
mal. PIDs are designed to detect rela-
tively low levels of gases.
The detector has become
saturated.
Move the Multi-PID 2 to a location
where it can sample clean air. Sample
clean air until the readings stabilize at
a low level.
Detector has been short-cir-
cuited by foreign matter in
the cell.
Contact Dräger Safety.
Situation
Cause
Action
Содержание Multi-PID 2
Страница 7: ...7 Notices and Warnings Notices and Warnings...
Страница 14: ...14 Notices and Warnings...
Страница 15: ...15 Introduction Introduction...
Страница 25: ...25 Using the Multi PID 2 Using the Multi PID 2...
Страница 42: ...42 Using the Multi PID 2...
Страница 43: ...43 Connecting Accessories Connecting Accessories...
Страница 52: ...52 Connecting Accessories...
Страница 53: ...53 Menu Functions Menu Functions...
Страница 71: ...71 Routine Maintenance Routine Maintenance...
Страница 86: ...86 Routine Maintenance...
Страница 87: ...87 Troubleshooting Troubleshooting...
Страница 93: ...93 Appendices Appendices...
Страница 104: ...104 Appendices...