61
Menu Functions
Figure 15 – Intervall time adjustment
Interval simultaneously displays the STEL, TWA, and REAL reading
along with the instantaneous reading. The Multi-PID 2’s Datalogger
can store up to 15000 entries. The interval you select will determine
the period of operation. At the end of the period, the datalogger will
be full and you will see the “Data Log Full” status.
Table 4 – Averaging Intervals and Period STEL, TWA, and REAL Operation
The Multi-PID 2’s display can show four values in Interval mode:
STEL, TWA, and REAL, plus the instantaneous reading.
Activating " Interval " mode:
1. Press the
" menu "
key and select
" Unit Setup "
, then press
" select "
. Select
" User Mode "
, then press
" select "
.
2. Select
" Interval "
, then press
" select "
Averaging Interval
(example)
Hours of Operation to
Fill the Datalogger
15 s
50
1 m
200
2 m
400
5 m
1000
15 m
3000
Multi-PID 2
ALARM
CHARGE
00323958_0
1
_en.eps
Intervals
Enter new
INTERVAL (sec)
0 0
1
5
✓
l
q
j
Содержание Multi-PID 2
Страница 7: ...7 Notices and Warnings Notices and Warnings...
Страница 14: ...14 Notices and Warnings...
Страница 15: ...15 Introduction Introduction...
Страница 25: ...25 Using the Multi PID 2 Using the Multi PID 2...
Страница 42: ...42 Using the Multi PID 2...
Страница 43: ...43 Connecting Accessories Connecting Accessories...
Страница 52: ...52 Connecting Accessories...
Страница 53: ...53 Menu Functions Menu Functions...
Страница 71: ...71 Routine Maintenance Routine Maintenance...
Страница 86: ...86 Routine Maintenance...
Страница 87: ...87 Troubleshooting Troubleshooting...
Страница 93: ...93 Appendices Appendices...
Страница 104: ...104 Appendices...