65
Menu Functions
Sample Collection
The
" Sample Collection "
function is used to suck a defined air vol-
ume through the device, e.g. to fill a preceding sampling tube (sam-
pling tube with activated carbon 67 33 011) or a gas-sampling tube
with a defined air volume. The sampling tube or the air collected using
the gas-sampling tube can be later analysed in a lab. This process is
menu-driven. The desired air volume in litres can be adjusted using
the arrow keys. It takes three minutes to take a sample of one litre.
In TVOC mode, the
" Sample Collection "
function is in the
" Unit
Setup "
menu. In Gas mode, this function is in the main menu.
1. Press the
" menu "
key. Highlight
" Unit Setup "
, then press
" select "
. Use the "
l
" key to highlight
" Sample Collection "
,
then press
" select "
.
2. Use the arrow keys "
q
" and "
l
" to set the value.
3. Press
" done "
to accept the set value.
Memory slots Menu
Only available in TVOC mode.
General Information
The response of a PID to a compound is dependent on its ionization
potential. In certain situations when the user is confident that only a
single specific compound exists, they can set up the Multi-PID 2
using an isobutylene calibration to mimic the Multi-PID 2 response as
if the Multi-PID 2 was calibrated using that specific single compound.
In situations where only a single pure compound is present in air, the
Multi-PID 2 should be calibrated with a standard of that specific com-
pound as span gas.
NOTE:
To utilize the » Mg/M3 « unit of measure, the instrument must be
configured and operated with a compound chosen from the
Multi-PID 2’s Preset RF library because a molecular weight value
is needed for the » Mg/M3 « calculation. If you wish to use the
» Mg/M3 « unit of measure while doing TVOC work, you should
use a memory slot with Isobutylene selected from the Preset RF
library. Refer to Section: Response Factors and Memory Slots
for further information.
Содержание Multi-PID 2
Страница 7: ...7 Notices and Warnings Notices and Warnings...
Страница 14: ...14 Notices and Warnings...
Страница 15: ...15 Introduction Introduction...
Страница 25: ...25 Using the Multi PID 2 Using the Multi PID 2...
Страница 42: ...42 Using the Multi PID 2...
Страница 43: ...43 Connecting Accessories Connecting Accessories...
Страница 52: ...52 Connecting Accessories...
Страница 53: ...53 Menu Functions Menu Functions...
Страница 71: ...71 Routine Maintenance Routine Maintenance...
Страница 86: ...86 Routine Maintenance...
Страница 87: ...87 Troubleshooting Troubleshooting...
Страница 93: ...93 Appendices Appendices...
Страница 104: ...104 Appendices...