75
Routine Maintenance
●
Before connection, ensure the gas cylinder valve and the regulator
CGA connection are clean.
●
Turn the pressure control valve on the cylinder all the way off (clo-
se the cylinder). Turn the regulator outlet to off. Open the gas cy-
linder valve slowly and check for leaks. Adjust the delivery
pressure and then open the regulator outlet valve.
Calibration Gas
Adequate ventilation must be provided when the Multi-PID 2 is being
calibrated.
If compound threshold limit values (TLV) are exceeded, you should
use a gas bag for sampling and calibration.
To determine the TLV of the compounds contained in the calibration
gas, refer to the Material Safety Data Sheet (MSDS) supplied with
your calibration gas cylinder.