RSR-000886-000 (5)
Page
15
of
20
Optical Specifications
Treatment Wavelengths
980 nm ± 20 nm & 810 nm ± 20 nm
Skin Tone Adjustments
Light (I, II) or Medium (III, IV)
Dark (V, VI)
Maximum Output Power
15W
15W
Wavelength Ratio
80% of 980 nm + 20% of 810 nm
100% 980 nm
Aiming Beam Output
<4.0 mW
Aiming Beam Wavelength
650 nm ± 20 nm
Operating Modes
Continuous Wave (CW) or Pulsed
Pulse Frequency
2, 10, 20, 100, 200, 500, 1000, 2500, 5000, 10000 Hz
Pulse Duration
50% duty cycle at any given pulse frequency setting
Laser Safety Goggles
OD5+ 808nm & 980nm
Electrical Specifications
Input Voltage
100-240 VAC; 50/60 Hz
Input Current
≤ 400 VA
Battery Pack
Li-Ion 9.9V, 5.0Ah
Beam Delivery Assembly (BDA) Specifications
BDA Model
New Empower Delivery System
Hand piece Assembly (-B6)
Basic hand piece (BHP) with 600μm fiber core
Empower Attachments
Flat Window-IQ, Small Cone-IQ, Small Massage Ball-IQ, Large Massage Ball-IQ, Large
Cone-IQ
(NOTE: “-IQ” indicates the attachment is the quick disconnect version with RFID tag)
7.2. Spare Parts
Warning:
Do not use any accessories and/or cables not specified or sold as replacement parts. Use of unauthorized accessories and/or
optical fibers not specified or sold for use with this system may result in unsafe treatment emissions or damage to the laser system.
Item
Description
Image
Beam Delivery
Assembly
Opaque protective double sheathed fiber with attached
Empower™ hand piece and attachments
Safety Goggles
808 nm and 980 nm OD 5+ Over glasses
Safety Goggles
808 nm and 980 nm OD 5+ Regular
Power Cord
Medical Grade Power Cord, 2.9 m
External Power Supply
Medical Grade Power Supply
Battery
49.5WHr, 9.9V, 5.0Ah (3S2P) Battery
7.3. Electromagnetic Compatibility (EMC) Tables
This laser system has been tested to comply with the requirements of EN/IEC 60601-1-2 for electromagnetic compatibility (EMC). The following
pages list the tests performed and corresponding test levels. The user, operator, installer, or assembler of this product is advised of the following:
•
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the
information provided in this document.
•
Portable and mobile RF communications equipment can affect medical electrical equipment.
•
Use only the power cord provided with this product, or an approved alternate.
Warning:
The use of accessories, transducers and cables other than those specified may result in increased emissions or decreased
immunity of this product.
Warning:
Equipment should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, this
device should be observed to verify normal operation in the configuration in which it will be used.
All medical electronic devices must conform to the requirements of EN/IEC 60601-1-2. Precautions, adherences to the EMC guideline information
provided, and verification of all medical devices in simultaneous operation are required to ensure the electromagnetic compatibility and coexistence
of all other medical devices prior to a laser therapy treatment.
7.3.1. Emissions Test Summary
Specification
Frequency Range
Compliance Status
CISPR 11/EN 55011
Group 1, A
Conducted Emissions
0.15 MHz – 30 MHz
PASS
CISPR 11/EN 55011
Group 1, A
Radiated Emissions
30.0 MHz – 1,000 MHz
PASS