6.3.1.2 Display in the force range setting mode
symbol for the
currently selected
parameter
here: setting the ex-
tension force range
set lower limit of
the force range for
random selection
here: 7 kg pressure
buttons to change settings
or confirm entries
here: up, down, right, OK
set upper limit of
the force range for
random selection
here: 14 kg pressure
Note
– Negative values (resistances) between -30 kg and -1 kg mean that the patient is required to pull
to overcome the set resistance, irrespective of the current direction of motion.
– Positive values (resistances) b1 kg and +30 kg mean that the patient is required to push
to overcome the set resistance, irrespective of the set direction of motion.
– If no resistance is set, the carriage will operate in the passive mode, moving in the direction of
extension or flexion.
– When you adjust the force range, the minimum value is shown on the left of the display and the
maximum value is shown on the right. The random values are selected from this range only.
– It is not possible to define a range that spans the value of 0 kg. The forces can be set to values
either greater than/equal to 0 or less than/equal to 0 in the respective direction of motion.
– If the set maximum value of a range equals the set minimum value, the patient always has to
exert this force; no random selection will be made.
– The random force value is only selected from the active range (see also: 5.3.2 ) and is only re-
quested for the duration of 4 seconds. Outside this range, the movement is always passive.