En
glis
h
67
The ARTROMOT®-K1 device is subject to par-
ticular precautions regarding electromagnetic
compatibility (EMC). The device must be installed
and put into service strictly in compliance with
the EMC directives set forth in the accompanying
documents.
Portable and mobile RF communication systems
may affect the ARTROMOT® ACTIVE-K.
The ARTROMOT® ACTIVE-K should not be used
adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, ARTROMOT®
ACTIVE-K should be observed to verify normal
operation in the configuration in which it will be
used.
If you detect damage or malfunctions that may
impair the safety of the patient or of the operator,
have the device repaired before using it.
If it is necessary to replace assemblies or cables,
only the manufacturer's original parts must be
used to ensure continued compliance with EMC
requirements after repair. This requirement ap-
plies to the power supply unit, cables and cable
lengths, drive unit consisting of the motor and
the control system, the programming unit incl.
the coiled cable and the connector.
The product designation ARTROMOT® ACTIVE-K
used in the text below includes all product
variants.
10. IEC 60601-1-2:2007 and IEC 60601-1-11:2010
10.1 Electromagnetic emissions
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
ARTROMOT® ACTIVE-K
is intended for use in the electromagnetic environment specified below. It
is the responsibility of the customer or user to ensure that the
ARTROMOT® ACTIVE-K
unit is used
in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions to CISPR 11
Group 1
ARTROMOT® ACTIVE-K uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions to CISPRH 11
Class B
ARTROMOT® ACTIVE-K is suitable for use in all
establishments, including domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions to
IEC 61000-3-2
complies with class
A requirements
Voltage fluctuations/flicker
emissions to IEC 61 000-3-3
complies
Supplied by EME Services Ltd