Fuse replacement
m
Warning!
Patient hazard, equipment malfunction and
damage —
The replacement of fuses must be referred
to specialists as defined in IEC 60364 or other
applicable standards (e.g. biomedical techni-
cians, electricians, electronics installers).
serial numbers < 2000
Fuses used must be T1A fuses.
serial numbers > 2000:
Fuses used must be T2A fuses.
Before replacing fuses, turn off the ARTRO-
MOT® ACTIVE-K and disconnect the device
from the power line.
Use an appropriate tool to remove the fuse
holder situated between the power switch
and the power connector (Fig. 1). Replace
the fuses and reinsert the fuse holder (Fig. 2).
Ensure that the fuse holder properly locks into
place.
Fig. 1
Fig. 2
7.3 Transport
The following operating steps must be com-
pleted before transporting the ARTROMOT®
ACTIVE-K :
1. Adjust the carriage to a femur length of
49 cm and a tibia length of 42 cm.
2. Activate the "transport setting" function
in the menu (see also 5.3) by selecting the
function, activating it withthe arrow up but-
ton and confirming the selection with OK.
3. The initial screen will automatically be dis-
played, allowing the carriage to be moved
to the transport position. Press START and
the carriage will automatically move to the
transport setting.
4. Turn off the ARTROMOT® ACTIVE-K power
switch (20).
5. Disconnect the power cord and the control
pendant.
6. Remove the footplate by fully opening
and unscrewing the clamping lever (14) for
adjustment of the footplate rotation.
7. Only use the original shipping box for trans-
porting the device. Ormed GmbH cannot
be held liable for transport damage if the
device is not packed in its original shipping
box.
8. Push the molded polystyrene blocks onto
the ARTROMOT® ACTIVE-K.
9. Place the clamping lever and the footplate
in their designated cut-outs in the lower
polystyrene board.
10. Store the power cord at the bottom of the
box, then place the ARTROMOT® ACTIVE-K
with attached polystyrene blocks in the box
as well.
11. Put the control pendant (22) in the supplied
cardboard container and store it in the box
of the ARTROMOT® ACTIVE-K.
footplate
clamping
lever to secure
footplate
control pendant
packed separately
power cord