En
glis
h
20
5. Setting the treatment values
m
Warning!
Patient hazard —
Before therapy
, a
test run
consisting of several exercise cycles must be
completed without the patient. Then repeat
the test run with the patient and check that
the movements do not cause any pain.
Note: See also 2.2 and 2.3!
5.1 General information
on programming
ARTROMOT® ACTIVE-K
Initial adjustment with chip card "Special"
(red)
Turn on the unit's power switch (20).
The display will show the main menu for selec-
tion of an operating mode (passive, active, coor-
dination), the therapy protocols, the documen-
tation function, the New Patient function, or the
Settings function. The active setting is indicated
with a green dot in the box below the symbol.
The header displays the
symbol, which rep-
resents the main menu. Next to it, you see the
current selection (here: passive (CPM)).
The scroll bar in the margin on the right shows
whether or not additional functions can be se-
lected. You select the options that are currently
out of sight by moving the focus (blue field) past
the top or bottom line of symbols.
The focus highlights the "passive" operating
mode.
You shift the focus by pressing the appropriate
arrow button.
Select one of the fields and activate the corre-
sponding setup menu with the MENU button.
The display will show the available therapy
parameters or sub-functions and the current
selections below the corresponding symbol.
You select a treatment parameter or function by
pressing the OK button while the focus is on that
particular field.
This is what happens when you press the OK
button to select a parameter:
− The corresponding symbol appears enlarged
on the display.
− The set value is displayed.
With the
arrow up/arrow down
buttons you
change the displayed value. When you press
and hold the button, the value will change at a
faster rate.
Some of the (special) functions can only be
enabled and disabled. This is done with the ar-
row up/arrow down buttons. Active parameters
are identified with a green square next to the
symbol and a check mark.
To confirm a selection, press the OK button.
Then the display returns to the higher-level
menu and you can continue making selections.
Next, press the
START
button: ARTROMOT®
ACTIVE-K automatically verifies the set values
and the sessions starts.
Supplied by EME Services Ltd