Note
– Preset parameters for all functions, unless otherwise stated:
– speed: 50 %
– active zone: 0° / 0° / 90° (for maximum setting, see 5.2.2).
– The "cool-down" function is the opposite of the "warm-up" function. After the session, the
carriage stops in the position equivalent to the set extension value + 10°.
– The indicated minutes are approximate values. Depending on the programmed maximum
range of motion and the set femur length, the times may vary.
– Normally the range of motion is set by means of the extension and flexion settings, outside
the therapy plans. If a session starts with the "free training" function, the final extension and
flexion values of this phase automatically become the ROM for the rest of the session.
– The operating modes and functions are described in detail in chapter: 6.
– The response threshold of the load reversal for the passive mode is set under "More” –
“Load Reversal”.
– During operation, the symbol representing the selected therapy plan and the actual level
are indicated in the header.
Supplied by EME Services Ltd