10.2 Electromagnetic immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
ARTROMOT® ACTIVE-K
is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the
ARTROMOT® ACTIVE-K
unit is used
in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic envi-
ronment - guidance
Electrostatic dis-
charge (ESD) to
IEC 61 000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast tran-
sient/bursts to
IEC 61 000-4-4
± 2 kV for
power supply lines
± 1 kV for input/output
lines
± 2 kV for
power supply lines
± 1 kV for input and
output lines
Mains power should be
that of a typical com-
mercial or hospital
environment.
Surges to
IEC 61 000-4-5
± 1 kV line-
to-line voltage
± 2 kV line-
to-earth voltage
± 1 kV line-
to-line voltage
± 2 kV line-
to-earth voltage
Mains power should be
that of a typical com-
mercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines to
IEC 61 000-4-11
< 5 %
U
T
(> 95 % dip in
U
T
) for
½ cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
< 5 %
U
T
(> 95 % dip in
U
T
) for 5 s
< 5 %
U
T
(> 95 % dip in
U
T
) for
½ cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
< 5 %
U
T
(> 95 % dip in
U
T
) for 5 s
Mains power should be
that of a typical com-
mercial or hospital
environment.
If the user of the
ARTROMOT® ACTIVE-K
unit requires continued
operation during power
mains interruptions, it is
recommended that the
ARTROMOT® ACTIVE-K
unit be powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field to
IEC 61 000-4-8
3 A/m
3 A/m
Power frequency mag-
netic fields should be at
levels characteristics of a
typical location in a typi-
cal commercial or hospital
environment.
NOTE
:
U
T
is the a.c. mains voltage prior to application of the test level.