16
Note: For an illustration of the individual
device components, see page 5.
4.1 Connecting the unit,
performance check,
Equipment supplied
The equipment supplied includes these items:
base unit, programming unit (control pendant,
22), patient chip card (23), power cord (not
shown), footplate (12), clamping lever to adjust
the footplate rotation and to fasten the foot-
plate (14), operating instructions
1. Mount the footplate (12) by slipping the
pin through the sensor housing (15) at the
foot bracket, screwing the clamping lever
(14) onto the end of the pin and locking the
clamping lever.
2. Connect the programming unit (22) to the
provided socket (17) by plugging it in and
screwing it tight.
3. Connect the
power cord
to socket (18) of
the unit and connect the
mains plug
to a
wall outlet with a non-fused earthed wire
(100 to 240 Volt, 50/60 Hz).
4. Turn the
power switch
(20) on.
5. Follow these steps to set the carriage to the
home position
.
–
Initial setup for new patients
Insert the
original patient chip card (23) into the
control pendant (22).
Press the
MENU
button
on the con-
trol pendant until you see the
main menu
(for selection of the operating mode).
Using the arrow buttons, move the focus
(blue field) for selection of a function (field
displayed with a blue frame) to highlight
the New Patient function
.
Select the function with
, then acti-
vate the function by pressing the arrow up
button
.
Confirm the selection with
. The ini-
tial screen for setting the unit to the home
position will be displayed.
Press
.
The unit automatically enters the home
position allowing you to perform the me-
chanical adjustment (40° flexion). On the
display, you will see the main menu.
– Adjustment with programmed chip
card
Insert the original patient chip card (23)
into the control pendant (22).
Press
.
The unit automatically enters the mount-
ing position (set ext 10°).
Performance check
If the control pendant can be operated as de-
scribed above and the
ARTROMOT® ACTIVE-K
enters the home position, the therapy unit has
passed the performance check.
The home position is:
− if a programmed chip card is used:
set extension angle + 10°
− after programming for New Patient:
stop position = 40 °
The device also runs performance checks regu-
larly during operation. This is what happens, if a
problem is identified:
− an audio signal sounds
− the unit switches off immediately
− the message ERROR and an error code
(e.g. ERROR 5) appear on the display.