Note
– Refer to sections 5.2, 5.3 and 6 for a descrip-
tion of the parameters.
– To
view
the
set parameters and the cur-
rently selected values
, press the MENU
button. However, this is only possible when
you press the STOP button fi rst.
– To prevent accidental changes of the param-
eter settings, you can
lock the buttons
. To
do so, simultaneously press the
arrow left/
arrow right buttons for approx. 4 seconds
while the carriage is stopped (all menus
closed): You will see the lockout menu that
allows you to disable to specifi c menu levels
(for details, see: section 5.3.2). Press both
buttons again for approx. 4 seconds to un-
lock and disable the lockout function in the
lockout menu.
– Selecting the New Patient (Reset) function
will automatically delete the data on the
patient chip card. When you have fi nished
programming the unit and press the
STOP
button, the settings will automatically also
be saved to the patient chip card.
–
Emergency stop function
: The ARTROMOT®
ACTIVE-K will stop immediately, when any
of the buttons is pressed during therapy. Pa-
tient treatment can be resumed by pressing
the START button. The unit will automatically
change the direction of movement.
–
Chip cards:
The explanations given for the
initial adjustment with a new patient are
based on the chip card “Special". With this
chip card, all available operating modes
and functions can be selected. To simplify
the use of the device, the display and
choice of operating modes and functions
can be limited. This is done by inserting dif-
ferent pre-programmed patient chip cards
on which certain functions are locked.
You can choose among the following
chip cards:
ACTIVE-K
passive
www.DJOglobal.com
Limited to
CPM functions
ACTIVE-K
www.DJOglobal.com
active
Limited to
CAM functions
ACTIVE-K
protocol
www.DJOglobal.com
Limited to
therapy plans
ACTIVE-K
special
www.DJOglobal.com
Free programming,
no limitations
If you use limited chip cards with locked
functions, the basic functions such as "Set-
tings”, “Load reversal” or “Documentation”
are accessed via the icon “More”
.
Patients with a programmed chip card
− Insert the chip card (the patient is not yet
positioned on the carriage).
− Press the
START
button: the carriage will
move to the middle position of the param-
eters stored on the chip card and stops.
− Perform the mechanical adjustments of the
unit (femur length, etc.).
− Position the patient on the carriage and press
the
START
button to initiate therapy.