C. PATIENT CONNECTIONS
1 - CO2 MAIN-STREAM MAINFLOW™ SENSOR
•
The MainFlow™ CO2 sensor is a rugged, solid-state, mainstream sensor. It is factory
calibrated and does not require further calibration.
2 - CO2 ADAPTERS
•
For monitoring CO2, select an airway adapter based on the patient and monitoring situation.
Airway adapter taper meets ISO 5356-1.
Airway Adapter Disposable for Main-Stream
MainFlow™ sensors
• Part Number CAP112
• For intubated patients with endotracheal tube
diameters greater than 4 mm
• Adds 5 cc of deadspace
• Weight: 7.7 grams
• Pressure drop: 0.40 cmH2O @ 60 LPM
• Color: Clear
Airway Adapter Disposable for Main-Stream
MainFlow™ sensors (Low Dead Space for
Neonatal)
• Part Number CAP113
• For intubated patients with endotracheal tube
diameters less than or equal to 4 mm
• Adds less than 1 cc of deadspace
• Weight: 9.1 grams
• Pressure drop: 0.74 cmH2O @ 10 LPM
• Color: Purple
Airway Adapter Reusable for Main-Stream
MainFlow™ sensors
• Part Number CAP110
• For intubated patients with endotracheal tube
diameters greater than 4 mm
• Adds 5 cc of deadspace
• Weight: 12.0 grams
• Pressure drop: 0.38 cmH2O @ 60 LPM
• Color: Black
Airway Adapter Reusable for Main-Stream
MainFlow™ sensors (Low Dead Space for
Neonatal)
• Catalog No. 7053
• For intubated patients with endotracheal tube
diameters less than or equal to 4 mm
• Adds less than 1 cc of deadspace
• Weight: 14.9 grams
• Pressure drop: 0.68 cmH2O @ 10 LPM
• Color: Red
Note
: All components are Latex free.
Note
: All dimensions are in millimeters.
Note
: Color pigment does not alter the specifications of the material.