Digicare Biomedical Technology
Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
8
46 – Patient Name and ID indicator
47- Date and Time indicators.
48 – Battery Status Indicator.
49 – Adult / Neonatal Mode selector and indicator.
50 – ECG Heart Rate (BPM) and ECG settings.
50 – Invasive Pressure 1 Indicator and IP1 settings.
51 – Invasive Pressure 2 Indicator and IP2 settings.
53 – SpO2 Indicator in % and SpO2 settings.
54 – Indicators and settings: End tidal CO2 (EtCO2) value display in %.
Inspired CO2 (InsCO2) value display in %.
Respiration Rate in RPM.
55 – FIO2 Indicator in % and settings:
56 – Indicators: NIBP mode selected indicator: Manual, Stat or Auto.
NIBP cycle interval selected indicator.
NIBP time to next cycle remaining indicator.
57 – Indicators: Systolic arterial pressure determined by NIBP in mmHg.
Diastolic arterial pressure determined by NIBP in mmHg.
Mean Arterial Pressure (MAP) determined by NIBP in mmHg.
58 – Temperature Indicators and settings
B. TURNING-ON
• Plug the AC power cord in AC MAINS receptacle at rear panel connector (24) and in the AC
outlet. The LifeWindow™ Lite LW8 accepts 110 or 220 VAC with automatic selection by the
power supply.
• The electrical installation of the room must have three pin outlets with earth connection for
protection of patient, users and equipment.
• The AC-ON LED indicator will light green and the BATTERY status indicator will light green if
the battery is fully charged or yellow if charging.
• Power-on the unit by double-clicking the ON switch ( 19 ) on the rear panel.
• The monitor will then perform a “boot” sequence (approx. 1.5 min.).
C. ALARM AND PULSE SOUND
• When the unit turns-on, the Alarm Sound is disabled by 2 minutes.
• To disable the Alarm Sound for more 2 minutes press and release the “Alarm Sound”
dedicated pushbutton ( 2 ) in the front panel. The AUDIBLE ALARMS OFF message flashes in
screen.
• To permanently disable the Alarm Sound press and hold 4 seconds the “Alarm Sound”
dedicated pushbutton ( 2 ). The AUDIBLE ALARMS OFF message is displayed steady on
screen.
• The Pulse and Alarm sound volume levels are set to default levels. To change the Pulse and
Alarm Sound volume level, touch in System menu and Sound Volume. Select the Beep
Volume, Alarm Volume and Beep Source, if it is from the ECG, SpO2 or OFF (no Beep).
D. PATIENT MANAGEMENT
• the LifeWindow™ Lite LW8 starts the monitoring screen and start storing all Vital Signs.
• The “Patient Management” screen can be opened at any time by touching the “System” menu.
• The “Patient Management” screen is automatically displayed when the monitor is turned-on
with two options: “Admit New Patient” and “Cancel Pat Mangmt”.
• Touching “Cancel Pat Mangmt”, closes the “Patient Management” screen.
• Although the Patient can be admitted at a later time, we recommend to admit the Patient
before Patient data starts to be stored to prevent losing Patient Data.