Digicare Biomedical Technology
Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
43
• The “Check Sample Line” message will clear and the sample pump will not be turned off if
the condition that caused the flow rate change is corrected within 2 minutes.
• The “Check Sample Line” message will clear if the sample cell is removed or if the
LifeWindow™ Lite LW8 is reset.
D. CO2 MONITORING
•
The LifeWindow™ Lite LW8 start displaying the CO2 waveform in two cascaded traces and
function and ETCO2, insCO2 and Respiration Rate measured values, with all alarm limits and
settings in the defaults values or the last saved settings.
1 – CO2 WAVEFORM AND SCALE
• The CO2 waveform traces are by default selected to traces 5 and 6 cascaded. The traces and
cascade options can be changed in the Traces menu. The traces settings return to the default
every time the unit is turned-on.
•
The CO2 waveform default sweep speed is 6.25 mm/sec. To change the CO2 waveform
sweep speed, touch the Traces menu and select the desired sweep speed. The options are:
6.25, 12.5, 25, 50 and 100 mm/sec
•
The CO2 waveform scale default is 0 to 50 mmHg. To change the CO2 waveform scale touch
the CO2 menu and select the desired waveform scale. The options are:
0 to 5 mm Hg, 0 to 10 mmHg, 0 to 50 mmHg and 0 to 100 mmHg.
2 – CO2 ALARM LIMITS
• CO2 Alarm Limits are set by default to:
ETCO2 Upper Limit: 45 mmHg
ETCO2 Lower Limit: 20 mmHg
InsCO2 Upper Limit: 10 mmHg
InsCO2 Lower Limit: OFF
Resp Rate Upper Limit: 50 BPM
Resp Rate Lower Limit: 5 BPM
• Change the CO2 Alarm Limits in the CO2 menu.
ETCO2 Upper Limit Range: 1 to 100 % – or OFF - non overlaping.
ETCO2 Lower Limit Range: 0 to 99 % – or OFF - non overlapping.
InsCO2 Upper Limit Range: 1 to 100 % – or OFF - non overlapping.
InsCO2 Lower Limit Range: 0 to 99 % – or OFF - non overlapping.
Resp Rate Upper Limit Range: 1 to 150 BPM – or OFF - non overlapping.
Resp Rate Lower Limit Range: 0 to 149 BPM – or OFF - non overlapping.
• To return to the factory default values, go to the SYSTEM menu and select the Load Factory
Default Settings.
3 – CO2 ALARM MESSAGES
High and Low ETCO
2
–
Are triggered if the ETCO
2
value overlaps the High or Low Limits
respectively.
High and Low InsCO
2
–
Are triggered if the InsCO
2
value overlaps the High or Low
Limits respectively.
High and Low Resp –
Are triggered if the Resp Rate value overlaps the High or Low
Limits respectively.
Sensor Faulty -
Indicates a hardware error in the module or module connection.
Reinsert or reset the module if necessary.