2 - FIO2 ALARM LIMITS
• The LifeWindow™ Lite LW8 FIO2 alarm limits default and range is:
FIO2 Upper Limit: OFF Range: 5 to 100%
FIO2 Lower Limit: 20% Range: 0 to 95%
• In the FIO2 menu, select the Upper and Lower alarm limits. To return to the factory DEFAULT
alarm limits, go to the SYSTEM menu and press the Load Factory Default Settings.
3 - FIO2 ALARM MESSAGES
FIO2 HIGH – Indicates the FIO2 reading is higher than the Upper Limit set.
FIO2 LOW – Indicates the FIO2 reading is lower than the Lower Limit set.
SECTION 11 – TECHNICAL SPECIFICATIONS
A. MECHANICAL DESCRIPTION
Size
: 9.25 x 6.89 x 6.49 inches (235 x 175 x 165 mm)
Weight
: 10lb (4.5Kg)
Color
: White and Gray
B. POWER REQUIREMENTS
AC voltage Input
: 100-240 VAC / 50-60 Hz max 80W
Universal Power Input
C. ECG
Lead Selection
: I, II, III (Standard lead configuration)
Input Isolation
: Isolated from Ground related circuits by
≥
4KV rms / 5.5 KV peak
Frequency Response
Filtered : 0.5 - 25Hz (
±
3db)
Unfiltered: 0.5 – 40Hz
Leakage Current
: < 10
µ
A at 120 to 240 Vac 50 / 60 Hz
Defibrillator & ESU Protection
: Max. 360J
Test Signal
: 1 mV / 100 ms at 70 BPM
±
2 BPM Pulse
Heart Rate
Range
: 15 to 300 BPM
Accuracy
:
±
1%
Resolution
: 1 BPM
Sensibility
Adult Mode
: 0.3 mV peak
Neonatal Mode
: 0.1 mV peak
QRS Detection
: Automatic in the selected mode
Pacemaker Rejection Amplitude
: 2mV to 700 mV
Duration
: 0.1 to 2 ms
Sweep Speed
: 6.25, 12.5, 25, 50 mm/sec
Lead Fault Alarm
: Audible and visual
Input
: 3 Lead ECG patient cable
Gain Selection
: Manual by operator
QRS Indicator
: Adjustable audible, visual
Patient Cable : 6 pin AAMI standard
CMRR :
≥
90 dB with patient cable and 51K
Ω
/ 47nF
imbalance
Input impedance :
≥
20 M
Ω
at 10 Hz with patient cable