Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
12
• Connect the temperature patient cable to receptacle (16 ) Temperature 1 or Temperature 2.
• In the Temperature menu select Temperature1, Temperature2 or T1-T2.
• The temperature value is displayed when the probe temperature is in the 50 to 122 °F (10 to
50°C), range otherwise the temperature display is blanked.
N. INVASIVE PRESSURE MONITORING
• The Invasive Pressure function is pre-configured with the following default settings. To change
the Invasive Pressure settings touch the IBPs menu and change the desired setting.
IP # : OFF
IP # Scale : 0 to +150 mmHg
IP # Mode : Pulsatile
IP # Label : IP #
IP # Trigger Source : Pulse
IP # Systolic Upper Limit : 160 mmHg
IP # Systolic Lower Limit : 60 mmHg
IP # Diastolic Upper Limit : 160 mmHg
IP # Diastolic Lower Limit : 60 mmHg
IP # Mean Upper Limit : 120 mmHg
IP # Mean Lower Limit : 60 mmHg
• The Invasive Pressure Channels 1 and 2 are turned off in the factory default settings, to avoid
alarms when not in use.
• Turn-on Invasive Pressure Channel 1 or 2 touching in the IBPs menu.
• The message “NO TRANSDUCER” is displayed until a Transducer is connected.
• Connect the pressure transducer cable to connector IP1 (14) or IP2 (18).
• Connect the catheter to the pressure transducer.
• Install the flow system and maintain the entire system with liquid.
• Hold the transducer at the heart level (axial line). Additional information in Section 9.
• The message “ZERO TRANSDUCER” is displayed until a zero calibration is executed.
• Open the TRANSDUCER input to the air (Local Atmospheric Pressure). Touch in the IBP
menu and touch in the Invasive Pressure 1 or 2. Touch the Zero Xducer button.
• If the transducer Zero calibration procedure was successful, the Invasive Pressure waveform
shall be displayed at ZERO baseline and the Invasive Pressure readings are displayed.
• If the transducer Zero calibration procedure was NOT successful, the “IP# ERROR
ZEROING” message is displayed on screen.
O. INSPIRED FRACTION OF OXYGEN MONITORING
• The FIO2 function is pre-configured with the following default settings. To change the FIO2
settings touch the FIO2 menu and change the setting.
FIO2 function : OFF
FIO2 Upper Limit : OFF
FIO2 Lower Limit : 20 %
• Turn-on the FIO2 function touching the FIO2 menu and select ON.
• Connect one end of the FIO2 patient cable into the jack receptacle on the back end of sensor.
• Connect the other end of the cable into the receptacle (15) located on the patient connections
panel of the unit.
• The LifeWindow™ Lite LW8 should be calibrated before each use and every 8 hours to
maintain maximum accuracy. See Section 10 for more information.
• The LifeWindow™ Lite LW8 can be used to measure a gas mixture for oxygen in two basic
modes:
• In the inhalation side of breathing circuit ahead of humidifiers and medicating devices or other
instances where gases are flowing to a patient in breathing circuits. When monitoring oxygen