Digicare Biomedical Technology
Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
10
• Selection of Patient Type shall be based on the Patient’s Heart Rate range for ECG monitoring
and Cuff size to be used for NIBP monitoring.
Adult Patient Type : ECG Heart Rate Range: 5 to 220 BPM
NIBP Cuff Size #: 5, 7, 8, 9, 10 and 11
Neonatal Patient Type : ECG Heart Rate Range: 30 to 350 BPM
NIBP Cuff Size #: 1, 2, 3, 4 and 6
I. ECG MONITORING
• The ECG function is pre-configured with the following default settings. To change the ECG
settings touch the ECG menu and change the desired setting.
Heart Rate High Limit : 140 BPM
Heart Rate Low Limit : 60 BPM
Heart Rate Averaging : 8 Beats for Adult, 16 Beats for Neonatal
ECG Scale : - 0.5 mV to +1.0 mV
ECG Filter Mains Notch : ON
ECG Filter Digital : ON
ECG Lead Selection : II
• Prepare the electrode sites in the standard configuration for RA (Right Arm), LA (Left Arm)
and LL (Left Leg). Application sites should be clean and dry. Shave or clip excess hair if
necessary. Use conductive ECG gel if necessary.
• Connect the 3 Leadwires to the ECG cable. Connect the ECG cable to the receptacle (13) in
the patient panel. Install the 3 electrodes in the patient. More details in Section 4.
• A visual indicator (
♥
) and a beep Sound indicates R wave detection and the ECG waveform
and Heart Rate reading are displayed.
J. SpO2 MONITORING
• The SpO2 function is pre-configured with the following default settings. To change the SpO2
settings touch the SpO2 menu and change the desired setting.
SpO2 % High Limit : OFF
SpO2 % Low Limit : 88%
SpO2 Pulse High Limit : 140 BPM
SpO2 Pulse Low Limit : 60 BPM
• Choose the SpO2 sensor type and prepare the SpO2 sensor and clip.
• The preferred sensor application site is the finger. Alternatively, the sensor and clip may be
applied to patient’s ear, hand or foot.
• Connect the sensor assembly to the SpO2 Patient Cable:
• Plug the SpO2 Patient Cable into the SpO2 connector (12) on the side panel of the monitor.
Push the cable in until you hear an audible “click”.
• The yellow SpO2 SENSOR message changes for the green Searching SpO2 message, the
Bar-graph shows the SpO2 Pulse Level, the SpO2 waveform, SpO2 % and Pulse Rate
values are displayed.
K. CO2 MONITORING
• The LifeWindow Lite LW8 automatically starts the CO2 function when CO2 sensor is
connected to the unit.
• The CO2 function is pre-configured with the following default settings. To change the CO2
settings touch the CO2 menu and change the desired setting.
CO2 Waveform Scale : 0 to 50 mmHg
EtCO2 High Limit : 45 mmHg