Digicare Biomedical Technology
Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
62
M. AUXILIARY OUTPUT
Pin 1 - Bio-Impedance Respiration analog waveform.
Pin 2 - ECGx1000 analog waveform.
Pin 3 - ECG Trigger Pulses: +12V positive pulses to defibrillator synchronism.
Pin 4 - Bio-Impedance Respiration trigger Pulses.
Pin 5 - Invasive Pressure 1 analog waveform.
Pin 6 - Invasive Pressure 2 analog waveform.
Pin 7 - NA
Pin 8 – NA
Pin 9, Pin10, Pin11, Pin12 - GND
Pin13 - RS232 – RXD optional depends on the model.
Pin14 - RS232 – TXD optional depends on the model.
Pin15 - RS232 – GND optional depends on the model.
DB15
SECTION 12 – MAINTENANCE
A. THE MONITOR
When necessary, clean the exterior surfaces of the monitor with a cloth or swab dampened with
warm and a mild detergent solution. Do not allow fluids to enter the interior of the instrument.
WARNING
: Electrical shock and flammability hazard - always turn the monitor off and
disconnect it from AC mains power before cleaning.
CAUTION:
Do not autoclave or pressure sterilize this monitor. Do not stack or immerse this
monitor in any liquid. Do not gas sterilize this monitor.
Do not touch, or rub the display panel with abrasive cleaning compounds, instruments,
brushes, rough surfaced materials or make contact with anything that can scratch the panel.
B. PROBES
•
The probes are the only surfaces of this monitor that come in contact with the patient. Clean
the probes after each patient use.
• Clean the monitor’s probes with a commercial cleaning solution before attaching a new
patient. Probes should be cleaned until signs of wear or splitting occur. At this time a new
probe is required.
• If disinfection is required, wipe the surfaces with Isopropyl alcohol or cider and use a water
rinse. When sterilization is required, use ethylene oxide and be sure to follow hospital
procedures.
• Inspect the probe for wear or splitting after every disinfecting/sterilization process is complete.
• If wearing or splitting of the probe is found upon visual inspection, a new probe should be
used
1 - MAINFLOW™ CO2 MAINSTREAM SENSOR
• Use a cloth dampened with isopropyl alcohol 70%, a 10% aqueous solution of sodium
hypochlorite (bleach), disinfectant spray cleaner such as Steris Coverage® Spray HB,
ammonia, or mild soap.
• Wipe down with a clean water-dampened cloth to rinse and dry before use. Make certain that
the sensor windows are clean and dry before reuse.
8 7 6
5
3 2
1
15
4
1 4 1 3 1 2 1 1 1 0 9