• Keep the MainFlow™ CO2 sensor attached to the adapter exposed to room air and away
from all sources of CO2, including the ventilator, the patient’s breath and your own.
• Touch in the CO2 menu and touch in the Calibration “Zero”. A Warning message: “Is input
exposed to 0 mmHg concentration” is displayed. Touch in the Yes button to perform a
Zero Calibration. The message Zero in Progress is displayed during the zero process,
typically 15 to 20 seconds (maximum time of 40 seconds).
5 – PATIENT CIRCUIT
•
Attach the MainFlow™ CO2 sensor and airway adapter to the patient circuit as close as
possible to the patient’s mouse.
D. CO2 MONITORING
•
The LifeWindow™ Lite LW8 start displaying the CO2 waveform in two cascaded traces and
function and ETCO2, insCO2 and Respiration Rate measured values, with all alarm limits and
settings in the defaults values or the last saved settings.
1 – CO2 WAVEFORM AND SCALE
• The CO2 waveform traces are by default selected to traces 5 and 6 cascaded. The traces and
cascade options can be changed in the TRACES menu. The traces settings return to the
default every time the unit is turned-on.
•
The CO2 waveform default sweep speed is 6.25 mm/sec. To change the CO2 waveform
sweep speed, touch the Traces menu and select the desired sweep speed. The options are:
6.25, 12.5, 25, 50 and 100 mm/sec
•
The CO2 waveform scale default is 0 to 50 mmHg. To change the CO2 waveform scale touch
the CO2 menu and select the desired waveform scale. The options are:
0 to 5 mm Hg, 0 to 10 mmHg, 0 to 50 mmHg and 0 to 100 mmHg.
2 – CO2 ALARM LIMITS
• CO2 Alarm Limits are set by default to:
ETCO2 Upper Limit: 45 mmHg
ETCO2 Lower Limit: 20 mmHg
InsCO2 Upper Limit: 10 mmHg
InsCO2 Lower Limit: OFF
Resp Rate Upper Limit: 50 BPM
Resp Rate Lower Limit: 5 BPM