OPERATION, CONT.
Body Mass Index (BMI) Operation with ID and DHR
Using Digital Height Rod
IMPORTANT! Before performing the operations on this page, the digital
height rod must be in its “starting position” (inner sliding tube must be
down completely inside outer stationary tube and headpiece folded flat
against stationary tube) prior to turning on the MV1.
1. Make sure height rod is in starting position.
2. Press the
ON/OFF
key to turn MV1 on.
3. If required, press the
ZERO
key to zero weight display. Note that height reading will
show
33’ 7.3”
or
110.0 cm
(depending on the
HEIGHT NITS
setting in Setup).
4. Perform tare operation if required. Refer to Push Button Tare or Tare Weight Entry in
previous section for instructions on using tare.
5. Press the
ID/HEIGHT
key.
6. Display will change to show
Id=
.
7. Using numeric keys, enter up to a 14-digit alpha numeric identification number.
8. Press the
ENTER
key.
NOTE:
If patient is shorter than 3' 7.3" (110 cm) or taller than 6' 7.3" (201.2 cm),
proceed to step 9, 10 or 11 depending on the
HEIGHT UNITS
setting. Otherwise,
press the
ENTER
key without entering a value for steps 9, 10 or 11 (twice for Step
9 and once for Steps 10 or 11) and proceed to Step 12.
9. If Centimeters was selected for
HEIGHT UNITS
in setup, the display will change to show
CENTIMETERS
.
a. Use numeric keys to enter up to 4 digits (# # # . #) for height in centimeters and
then press the
ENTER
key.
10. If Millimeters was selected for
HEIGHT UNITS
in setup, display will change to show
MILLIMETERS
.
a. Use numeric keys to enter up to 4 digits (# # # #) for height in millimeters and then
press the
ENTER
key.
11. Assist patient onto scale.
12. Grasp height rod at hinge pin (see Figure No. 10)
and raise it well above patient’s head, and then lift
headpiece to horizontal position. Carefully lower
height rod until headpiece rests on patient’s head.
13. BMI will automatically be displayed with height.
14. When weight is stable (STABLE
[\
annunciator will
turn on), read weight, height, and BMI displays.
15. If a printer is connected to scale, press the
key to print a ticket.
16. While holding headpiece horizontal, raise height rod
well above patient’s head.
17. Assist patient off scale and return height rod to
starting position.
18. If Tare operation was performed, press the
NET/GROSS
key to return to Gross weight mode.
19. Press the
ZERO
key. This will reset tare weight to zero.
20. BMI display and ID will clear when weight returns to zero.
Figure No. 10
Содержание MedVue MV1
Страница 1: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 1 Model MV1 Medical Weight Analyzer Owner s Manual ...
Страница 2: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 2 ...
Страница 6: ......
Страница 57: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 51 PARTS IDENTIFICATION CONT ...
Страница 61: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 55 ...