8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
37
CALIBRATION SETUP, CONT.
SSERIAL PORT
With
SERIAL PORT
displayed, press the
ENTER
key. The display will change to
RS232 MOdE
.
Proceed to the
RS232 MOdE
(RS232 Port Output Mode) parameter.
Otherwise, press the
ID/HEIGHT
key to advance to the
USb PORT
menu.
RS232 MOdE
(RS232 Port Output Mode)
`
The display will show
RS232 MOdE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0, 1, 2, or 3.
0 = RS232 Port Output Mode is Disabled
1 = RS232 Port Output Mode is set for Weight on Demand
2 = RS232 Port Output Mode is set for Continuous Data
3 = RS232 Port Output Mode is set for a Printer
bAud RATE
(Serial Output Baud Rate –
Only if RS232
MOdE = 1, 2
or
3
)
The display will show
bAud RATE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 1 through 9, and 0.
1 = 2400 Baud
4 = 14.4k Baud
7 = 38.4k Baud
2 = 4800 Baud
5 = 19.2k Baud
8 = 56k Baud
3 = 9600 Baud
6 = 28.8k Baud
9 = 57.6k Baud
0 = 115.2k Baud
PRINTER TYPE
(Set Type of Printer Output –
Only if RS232
MOdE = 3
)
`
The display will show
PRINTER TYPE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0, 1, 2, or 3.
0 = Print default ticket (P185 or other type of ASCII printer)
1 = Print journal line
2 = Print label (P220)
3 = Print nControl ticket (must have loaded a ticket format, otherwise will use
default ticket)
ENd OF PRINT
(End of Print Linefeeds –
Only if RS232
MOdE = 3
)
At the end of data sent to a printer, the MV1 can send a number of carriage return/line feeds
to space the paper in the printer to a desired position for withdrawal or for the next print.
The display will show
End OF PRINT
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0 through 9.
Содержание MedVue MV1
Страница 1: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 1 Model MV1 Medical Weight Analyzer Owner s Manual ...
Страница 2: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 2 ...
Страница 6: ......
Страница 57: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 51 PARTS IDENTIFICATION CONT ...
Страница 61: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 55 ...