8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
34
CALIBRATION SETUP, CONT.
SSLEEP MOdE
(Sleep Mode)
The Sleep Mode feature conserves power when the MV1 remains unused for a selected
period of time (approximately 1 to 9 minutes). When enabled, the display will show the date
and time and the backlight will be turned off. The Sleep Mode feature requires the MV1 to
remain at a stable zero weight to activate. Weight placed on the scale or pressing any key
will activate the MV1 and return it to weight mode.
The display will show
SLEEP MOdE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0 through 9.
NOTE:
Selecting 0 disables the Sleep Mode feature.
AUTO SHUT OFF
(Auto-Shut Off Feature)
The Auto Shut Off feature will automatically turn the MV1 off (when it is not in use) after a
predetermined period of inactivity (approximately 1 to 9 minutes) to conserve power. To turn
the MV1 back on you must press the
ON/OFF
key.
The display will show
AUTO SHUT OFF
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0 through 9.
NOTE:
Selecting 0 disables the Auto Shut Off feature.
KEY dISAbLE
(Disable Key Operation)
This parameter determines which keys (
NET/GROSS
,
TARE
, and/or
ID/HEIGHT
) are
locked out during normal operation.
The display will show
KEY dISAbLE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0, 1, 2, or 3.
0 = No Keys Disabled
1 = NET/GROSS and TARE Keys Disabled
2 = ID/HEIGHT Keys Disabled
3 = NET/GROSS, TARE, and ID/HEIGHT Keys Disabled
ENAbLE Id
(ID Number Feature)
This setting determines whether the MV1 operates in the Transaction Storage Mode (no ID
storage) or in the ID Storage Mode and whether the MV1 allows an ID number to be entered
prior to beginning the weighing operation. This applies to both Basic Operation and with the
BMI feature enabled.
NOTE:
The ID Storage Mode allows up to 99 ID’s.
The display will show
ENAbLE Id
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise, use the numeric keys 0/NO or
1/YES
to select a new setting, and then press the
ENTER
key to save it.
Allowable values are YES or no.
YES = ID Enabled, uses ID
Storage Mode
no = ID Disabled, uses Transaction
Storage Mode
Содержание MedVue MV1
Страница 1: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 1 Model MV1 Medical Weight Analyzer Owner s Manual ...
Страница 2: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 2 ...
Страница 6: ......
Страница 57: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 51 PARTS IDENTIFICATION CONT ...
Страница 61: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 55 ...