8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
26
CALIBRATION SETUP, CONT.
ZZERO
With
ZERO
displayed, press the
ENTER
key. The display will change to
ZERO LIMIT
. Proceed
to the
ZERO LIMIT
(Two Percent Zero Tracking Range Limit) parameter.
Otherwise, press the
ID/HEIGHT
key to advance to the
FILTER
menu.
ZERO LIMIT
(Two Percent Zero Tracking Range Limit)
This setting limits the scale to zero only + or - 2% of the scale capacity.
If enabled, (
ZERO LIMIT = YES
) zero tracking and push button zero is limited to 2% of
scale capacity.
If disabled, (
ZERO LIMIT = no
) zero tracking and push button zero will operate up to full
scale capacity.
The display will show
ZERO LIMIT
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise, use the numeric keys 0/NO or
1/YES
to select a new setting, and then press the
ENTER
key to save it.
Allowable values are YES or no.
YES = 2% Zero Limit is Enabled
no = 2% Zero Limit is Disabled
POWER UP ZERO
(Power Up Zero)
The display will show
POWER UP ZERO
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise, use the numeric keys 0/NO or
1/YES
to select a new setting, and then press the
ENTER
key to save it.
Allowable values are YES or no.
YES = Weight display will be reset to
zero automatically on power up
no = Weight display will not be reset to
zero on power up
ZERO TRACKING
(Zero Tracking)
The zero tracking is a value in scale divisions that will automatically zero the scale within the
limits determined by the setting for this prompt.
The display will show
ZERO TRACkING
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0 through 18 (1 to 9 divisions by 0.5 divisions).
For example, 1 = .5 grads, 2 = 1 grad, 3 = 1.5 grads, etc...
NOTE:
Entering 0 (zero) will disable zero tracking.
Содержание MedVue MV1
Страница 1: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 1 Model MV1 Medical Weight Analyzer Owner s Manual ...
Страница 2: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 2 ...
Страница 6: ......
Страница 57: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 51 PARTS IDENTIFICATION CONT ...
Страница 61: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 55 ...