8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
31
CALIBRATION SETUP, CONT.
GGRAVITY
With
GRAVITY
displayed, press the
ENTER
key. The display will change to
GRAVITY ENAbLE
.
Proceed to the
GRAVITY ENAbLE
(Enable Gravity Constant) parameter.
Otherwise, press the
ID/HEIGHT
key to advance to the
LOCK
menu.
GRAVITY ENAbLE
(Enable Gravity Constant)
The Detecto MedVue
®
Model MV1 Weight Analyzer is equipped with a gravity constant
function which means that it can be calibrated in one location and then adjusted to match
the gravity at the location where it will used.
The display will show
GRAVITY ENAbLE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise, use the numeric keys 0/NO or
1/YES
to select a new setting, and then press the
ENTER
key to save it.
Allowable values are YES or no.
YES = Enable Gravity Constant and
proceed to
CAL GRAVITY
parameter
no = Disable Gravity Constant and
proceed to the
HOLd MOdE
parameter
CAL GRAVITY
(Calibrated Gravity Constant –
Only if
GRAVITY ENAbLE = YES
)
This is the gravity constant of the location where the scale was calibrated.
NOTE:
Consult the factory Tech Support for the Gravity Constant for your location.
The display will show
CAL GRAVITY
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a
new value, and then press the
ENTER
key to save it.
Allowable values are 0 through 99,999.
OP GRAVITY
(Operating Gravity Constant –
Only if
GRAVITY ENAbLE = YES
)
This is the gravity constant for the location where the scale will be operated.
NOTE:
Consult the factory Tech Support for the Gravity Constant for your location.
The display will show
OP GRAVITY
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a
new value, and then press the
ENTER
key to save it.
Allowable values are 0 through 99,999.
Содержание MedVue MV1
Страница 1: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 1 Model MV1 Medical Weight Analyzer Owner s Manual ...
Страница 2: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 2 ...
Страница 6: ......
Страница 57: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 51 PARTS IDENTIFICATION CONT ...
Страница 61: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 55 ...