8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
28
CALIBRATION SETUP, CONT.
CCALIbRATION
With
CALIbRATION
displayed, press the
ENTER
key. The display will change to
CALIbRATE
.
Proceed to the
CALIbRATE
(Perform Calibration) parameter.
Otherwise, press the
ID/HEIGHT
key to advance to the
GRAVITY
menu.
CALIbRATE
(Perform Calibration)
The display will show
CALIbRATE
and the current setting (no).
x
If the scale has been previously calibrated and you wish to skip calibration (and retain
the internal calibration factor), press the
ENTER
key. The display will change to
GRAVITY ENAbLE
. Proceed to the
GRAVITY
menu,
GRAVITY ENAbLE
parameter.
x
If calibration is desired, press the
1/YES
key to select YES, and then press the
ENTER
key. The display will change to
NUM CAL POINTS
.
Allowable values are YES or no.
YES = Perform Calibration and
proceed to
NUM CAL POINTS
no = Skip Calibration and proceed to
the
GRAVITY ENAbLE
parameter
NUM CAL POINTS
(Number of Calibration Points –
Only if
CALIbRATE
= YES
)
The MedVue Model MV1 Medical Weight Analyzer has two (2) modes that can be used to
perform calibration. The calibration modes are as follows:
Single Calibration Point
This is a standard calibration method requiring one weight, an empty scale, and has one
conversion factor. This method uses two calibration points (a test load on the scale and an
empty scale) to establish a zero (no load) calibration value and to span the MV2.
Multi-point Calibration
This method requires two weights, an empty scale, and has two conversion factors. It uses
three calibration points (a load on the scale, half the previous load and an empty scale).
The three points correspond to zero weight, mid-point weight, and test load (weight).
The display will show
NUM CAL POINTS
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 1 or 2.
1 = Single Calibration Point
2 = Multi-Point Calibration
Содержание MedVue MV1
Страница 1: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 1 Model MV1 Medical Weight Analyzer Owner s Manual ...
Страница 2: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 2 ...
Страница 6: ......
Страница 57: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 51 PARTS IDENTIFICATION CONT ...
Страница 61: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 55 ...