MedVue MV1 Owner’s
47
ERROR AND STATUS DISPLAYS
Display
Meaning
ERROR
General error, invalid keypad entry was attempted.
ERR OVERFLOW
Attempting to display a negative number greater than –9,999 or a
positive number greater than 99,999
ERR ZERO LIM
Indicates an attempt to zero a weight outside scale zero range.
Refer to the Calibration Setup
ZZERO
Menu,
ZERO LIMIT
(Two
Percent Zero Tracking Range Limit) parameter.
ERR UNSTAbLE
Motion is present when MV1 is attempting to perform one of the
following operations: Power Up Zero or Zero Weight Display
OVER CAPACITY
Scale weight exceeds scale capacity.
`
When the batteries are near the point they need to be replaced
(Alkaline) or recharged (NiCad or NiMH), the battery status
symbol will not show any bars. If the battery voltage drops too low
for accurate weighing, the MV1 will automatically shut off and you
will be unable to turn it back on.
CAL REQUIREd
Indicates improper stored calibration data. Calibration of MV1 is
required. Consult your scale service representative.
ERR ANALOG
An analog to digital convertor communication error has been
detected. Consult your scale service representative.
ANALOG LOW
The load cell input is below the range of the MV1. Consult your
scale service representative.
ANALOG HI
The load cell input is above the range of the MV1. Consult your
scale service representative.
ERR EEPROM
NOVRAM failure. Consult your scale service representative.
BEFORE YOU CALL FOR SERVICE
Problem
Possible Solutions
Display does not turn on AC Operation:
x
Is AC power supply fully inserted into wall receptacle?
x
Check wall receptacle for proper AC power. Try another
electrical appliance in same receptacle, does it work?
x
Check circuit breaker.
x
Has there been power failure?
Battery operation:
x
Check if batteries are installed and correctly.
x
Is the battery status symbol empty (no bars are displayed)?
If Alkaline, remove old batteries and replace with new ones.
If NI-CAD or NiMH, remove discharged batteries and
replace with fully charged ones. Place discharged batteries
in an external charger to recharge.
Incorrect weight is
displayed
Ensure that scale platform isn't touching an adjacent object. Have
proper operation procedures been followed?
Weight is not displayed
Refer to Error and Status Display section.
Height is not displayed
The digital height rod was not in its “starting position” before
turning on the MV1. The inner sliding tube must be completely
down inside outer stationary tube and headpiece folded flat
against stationary tube prior to turning on the MV1.
Содержание MedVue MV1
Страница 1: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 1 Model MV1 Medical Weight Analyzer Owner s Manual ...
Страница 2: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 2 ...
Страница 6: ......
Страница 57: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 51 PARTS IDENTIFICATION CONT ...
Страница 61: ...8555 M512 O1 Rev E y MedVue MV1 Owner s 55 ...