ENGLISH • 12
• If sensitivity persists, decrease power setting and/or move
from the sensitive tooth to another and then return.
Section 10: System Care
10.1
Daily Protocol
Start-up procedures at the beginning of the day
1. Switch ON the dental treatment unit (refer to instruction manual
of treatment unit).
2. Remove the DENTSPLY Cavitron
®
cable connector assembly
from handpiece holder and set power adjustment to 30% of
maximum.
3. Connect Steri-Mate
®
handpiece to the cable connector assembly
(refer to the Infection Control information booklet for sterilizing
instructions).
4. Adjust the lavage control to maximum.
5. Hold the handpiece (without an insert installed) over a sink
or drain. Activate the foot control and flush water through
handpiece for at least two minutes.
6.
Place a sterilized insert into the handpiece using a gentle push-
twist motion. Refer to the Cavitron Ultrasonic Inserts Directions
for Use for cleaning and sterilizing instructions.
7. Activate the ultrasonics and adjust the power and lavage control
to your preferred operating positions.
accidental prolonged contact with the activated tip
whenever the insert is placed in the mouth.
• Always adjust the lavage so adequate fluid will be
available to cool the tip-tooth interface.
• In general, it is suggested that a “feather light touch” be
used both supra- and subgingivally. The motion of the
activated tip and acoustic effects of the irrigating fluid,
in most cases, is adequate to remove even the most
tenacious calculus.
• Set the System’s power adjustment knob to the lowest
power setting for the application and the selected insert.
• Only adjust the System’s power adjustment knob with
the insert outside the patients mouth.
9.3 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. Point should be directed away
from root surfaces.
• Not keeping the tip in motion on the tooth. Do not allow
the insert to remain in a static position on any one area
of the tooth. Change the insert’s path of motion.
• Applying pressure. Use extremely light grasp and
pressure, especially on exposed cementum.
Содержание Cavitron SPS G90
Страница 2: ......
Страница 17: ...ENGLISH 15...
Страница 18: ...ENGLISH 16...
Страница 19: ...ENGLISH 17...
Страница 20: ...ENGLISH 18 Optional...
Страница 36: ...FRAN AIS 34...
Страница 37: ...FRAN AIS 35...
Страница 38: ...FRAN AIS 36...
Страница 39: ...FRAN AIS 37 En option...
Страница 55: ...ESPA OL 53...
Страница 56: ...ESPA OL 54...
Страница 57: ...ESPA OL 55...
Страница 58: ...ESPA OL 56 Opcional...
Страница 74: ...DEUTSCH 72...
Страница 75: ...DEUTSCH 73...
Страница 76: ...DEUTSCH 74...
Страница 77: ...DEUTSCH 75 Wahlweise...
Страница 93: ...ITALIANO 91...
Страница 94: ...ITALIANO 92...
Страница 95: ...ITALIANO 93...
Страница 96: ...ITALIANO 94 Facoltativa...
Страница 101: ...99 4 5 9 Sterile Lavage Kit P N 81340 5 0 12...
Страница 105: ...103 6 2 25 172 60 414 60 Cavitron 25 C 77 F Cavitron 7 SPS SPS 7 1 DENTSPLY Cavitron SPS...
Страница 106: ...104 7 2 Steri Mate Cavitron 30K Steri Mate...
Страница 107: ...105 DENTSPLY 7346 30K FSI SLI 10S 7 3 DENTSPLY Cavitron 30K DENTSPLY Cavitron...
Страница 108: ...106 8 Steri Mate DENTSPLY Cavitron 9 9 1 High Vacuum Evacuator HVE 9 2 Cavitron 7 4...
Страница 109: ...107 Blue Zone 9 3 10 10 1 1 2 DENTSPLY Cavitron 30 3 Steri Mate 4 5 6 Cavitron 7 1 Steri Mate Cavitron...
Страница 110: ...108 2 3 4 Steri Mate 5 30 6 1 Cavitron 2 Steri Mate Cavitron 3 4 5 6...
Страница 111: ...109 11...
Страница 112: ...110...
Страница 113: ...111...
Страница 114: ...112...
Страница 115: ...113...
Страница 117: ......