OXITEST PLUS 7 OPERATING MANUAL
3.12 Saving and Printing Test Results
Up to 20 PASS/FAIL results from manually-conducted tests may be saved in the memory of
Oxitest Plus 7
for later printout or downloading to a personal computer (PC). In addition to the 20
manual test results,
Oxitest Plus 7
can also save the results of Auto Preset tests and Alarm Tests.
Manual test and Auto Preset test results are saved to
Oxitest Plus 7
memory by pressing either the
Pass
or
Fail
key on the front panel keypad. The information saved includes the test SpO
2
, heart
rate, pulse amplitude and motion artifact settings, plus the PASS or FAIL status of the test and the
oximeter signal status displayed at the upper right corner of the
Oxitest Plus 7
LCD. The oximeter
signal status is saved in abbreviated format as follows: OK = “RED+IR OK”; NoRD = “No Red”;
NoIR = “ No IR”; and ?? = “NO SIGNAL” or “Testing”.
For each Alarm Test performed, the oximeter ALARM-ON and ALARM-OFF times are
automatically saved in the
Oxitest Plus 7
memory if the Alarm Test is run through to completion.
There is no need to press
Pass
or
Fail
to save the results of an Alarm Test.
All test results saved in the
Oxitest Plus 7
memory are retained by battery back-up when the power
is switched off. This allows test results to be acquired over a period of time and then printed out or
downloaded later.
Pressing the
key causes the test results saved in the memory to be transmitted from the
Oxitest Plus 7
RS-232 Port as a formatted “Test Report”. The
Oxitest Plus 7
RS-232 Port may
be connected to a Citizen iDP-3110 serial printer or to a personal computer. Refer to section 4.2 for
the interconnection cable details.
The format of the Test Report is similar to the following example printout:
Operation/Chapter 3
#
Page 22