OXITEST PLUS 7 OPERATING MANUAL
Pressing the
Alarm Test
key produces the following LCD display:
LOW SAT ALM T: 00.0
97% 72 BPM 100%
As shown above, the LCD displays the name of the alarm test at the upper left, a timer counter at
the upper right, and the test settings corresponding to the "normal patient" condition on the lower
line.
By pressing the
Alarm Test
key, the user can select one of the four remaining alarm tests. In
addition to the test above, the remaining tests appear as follows:
HI RATE ALM T: 00.0
97% 72 BPM 100%
LO RATE ALM T: 00.0
97% 72 BPM 100%
LOW SIG ALM T: 00.0
97% 72 BPM 100%
ARTFACT ALM T: 00.0
97% 72 BPM 100%
As long as the user refrains from pressing the
Start/Stop
key,
Oxitest Plus 7
continuously outputs
the simulation corresponding to the parameters shown on the lower line of the LCD. Pressing the
Start/Stop
key immediately initiates the selected alarm test by switching the test parameters to the
abnormal patient setting, and starts the timer counting in the top right corner of the LCD.
Operation/Chapter 3
#
Page 19