OXITEST PLUS 7 OPERATING MANUAL
1.7
Alarm Tests
Five automated test sequences for determining oximeter alarm response time to Low SpO
2
;
Low Heart Rate; High Heart Rate; Low Perfusion; and Motion Artifact conditions. Also
determines time required for oximeter alarm to self-clear after test conditions are returned to
normal.
Alarm Test parameters may be user-programmed via the RS-232 port.
1.8
Memory
Auto Preset and Alarm Test results saved in battery-backed, non-volatile Memory, along
with up to 20 PASS/FAIL records from manually-conducted tests. Test results in Memory
may be printed as a Test Report via optional external Printer, or may be downloaded via RS-
232 port to personal computer or automated tester.
1.9
Pulse Oximeter Make / Model
Major makes and models supported. Refer to the
Oxitest Plus 7
product CD, or
www.datrend.com,
for a detailed listing.
1.10
User Interface
Display:
LCD - 20 character x 2 line
Keypad - 15 keys:
PULSE OX, SPO
2
(%), HEART RATE (preset), HR
•
, HR
–
, PULSE AMP (preset),
PUL
•
, PUL
–
, PASS, FAIL, AUTO (preset), ALARM TEST, START/STOP (alarm
test), ERASE (memory), PRINT
1.11
Nellcor Port
Allows temporary connection of Nellcor pulse oximeter sensors for selection of sensor-
specific SpO
2
calibration curve. Intended for high-accuracy testing of all Nellcor disposable
and re-usable sensors.
Specifications/Chapter 1
#